    Use of multisite quantitative ultrasonography for noninvasive assessment of bone in horses (2002)

    Art
    Zeitschriftenartikel / wissenschaftlicher Beitrag
    Autoren
    Carstanjen, Bianca
    Lepage, Olivier M
    Detilleux, Johann
    Duboeuf, François
    Amory, Hélène
    Quelle
    American journal of veterinary research : AJVR
    Bandzählung: 63
    Heftzählung: 10
    Seiten: 1464 – 1469
    ISSN: 0002-9645
    Sprache
    Englisch
    Verweise
    Pubmed: 12371776
    Kontakt
    Pferdeklinik

    Oertzenweg 19 b
    14163 Berlin
    +49 30 838 62299 / 62300
    pferdeklinik@vetmed.fu-berlin.de

    Abstract / Zusammenfassung

    To evaluate the usefulness of multisite quantitative ultrasonography for noninvasive assessment of bone in horses.

    12 healthy horses and both forelimbs from 8 clinically normal horses.

    For in vivo measurements, various regions of interest (ROI) were examined on the third metacarpal bone, radius, and tibia. Precision error for speed of sound (SOS) measurements was obtained by measuring each ROI of 4 horses 10 times with probe repositioning. Additionally, 3 operators measured each aspect of the third metacarpal bone of 6 horses 5 times each. For ex vivo measurements, third metacarpal bones were examined at 9 ROI, and SOS measurements were performed before and after soft tissue removal. One ROI of a single forelimb was subjected to 96 ex vivo measurements with 3 different contact media.

    The lateral aspect of the third metacarpal bone had significantly higher SOS values than the dorsal and medial aspect of the third metacarpal bone. No difference was obtained between SOS values of the lateral and medial aspect of the radius. The tibia had significantly higher SOS values than the lateral aspect of the radius and the dorsal and medial aspect of the third metacarpal bone. Intraoperator coefficients of variation ranged from 0.62 to 3.15%, and interoperator coefficients of variation ranged from 0.78 to 2.70%. Values of SOS were highest when silicone oil was used as the contact medium.

    Speed of sound measurements obtained by quantitative ultrasonography in axial transmission mode can be used to precisely measure superficial cortical bone properties of third metacarpal bone, radius, and tibia in horses.