    Clinical evaluation of the CA530-VET hematology analyzer for use in veterinary practice (2007)

    Art
    Zeitschriftenartikel / wissenschaftlicher Beitrag
    Autoren
    Roleff, Sabine
    Arndt, Gisela
    Bottema, Bram
    Junker, Lothar
    Grabner, Arthur
    Kohn, Barbara
    Quelle
    Veterinary clinical pathology
    Bandzählung: 36
    Heftzählung: 2
    Seiten: 155 – 166
    ISSN: 0275-6382
    Sprache
    Englisch
    Verweise
    Pubmed: 17523089
    Kontakt
    Pferdeklinik

    Oertzenweg 19 b
    14163 Berlin
    +49 30 838 62299 / 62300
    pferdeklinik@vetmed.fu-berlin.de

    Abstract / Zusammenfassung

    The CA530-VET is a completely automated impedance cell hematology analyzer, which yields a 16-parameter blood count including a 3-part leukocyte differential.

    The aim of this study was to examine the operational potential of the CA530-VET and its value for use in veterinary practice.

    The analyzer was tested for blood carry-over, precision, and accuracy. Comparison methods included the CELL-DYN 3500, microhematocrit centrifugation, manual platelet (PLT) counting for feline and equine species, and a 100-cell manual WBC differential. Blood samples for comparison of the methods were obtained from 242 dogs, 166 cats, and 144 horses.

    The carry-over ratio (K) was 0.28% for RBC, 0.59% for PLT, 0.32% for WBC, and 0.18% for hemoglobin (HGB) concentration. Coefficients of variation (CVs) for within-batch precision and duplicate measurement of blood samples were clearly within the required limits, except for duplicate platelet counts in cats (8.7%) and horses (9.5%). The WBC count was in excellent agreement for dogs and horses and RBC count was in excellent agreement for horses. The accuracy of feline WBC counts was not acceptable, with the exception of values at the high end of the range. RBC counts in dogs and cats, and HGB concentration and MCV in all 3 species were sufficiently accurate. The CA530-VET HCT results were in excellent agreement with microhematocrit results in horses but exceeded the maximum allowed inaccuracy for cats and dogs. In all species, PLT counts established mechanically and manually were not in adequate agreement. Large differences were found between the CA530-VET and the manual differential percentage for lymphocytes and "mid-sized cells" (monocytes and basophilic granulocytes).

    The CA530-VET can be considered useful for routine canine, feline, and equine blood cell analyses. It should not be considered accurate, however, for PLT counts, feline total WBC counts in the subnormal and normal range, and leukocyte differentials, except for granulocytes.