    Validation of a portable hand-held whole-blood ketone meter for use in cats (2012)

    Art
    Zeitschriftenartikel / wissenschaftlicher Beitrag
    Autoren
    Weingart, Christiane (WE 20)
    Lotz, Fabian (WE 16)
    Kohn, Barbara (WE 20)
    Quelle
    Veterinary clinical pathology
    Bandzählung: 41
    Heftzählung: 1
    Seiten: 114 – 118
    ISSN: 0275-6382
    Sprache
    Englisch
    Verweise
    DOI: 10.1111/j.1939-165X.2011.00389.x
    Pubmed: 22250845
    Kontakt
    Klein- und Heimtierklinik

    Oertzenweg 19 b
    14163 Berlin
    +49 30 838 62422
    kleintierklinik@vetmed.fu-berlin.de

    Abstract / Zusammenfassung

    Urinary dipsticks are the most frequent method used for screening of ketones in animals, but this method has many drawbacks. In human medicine, portable meters that measure ketones in whole blood have largely replaced urinary dipsticks.

    The aim of this prospective study was to validate a portable whole-blood ketone meter for use in cats.

    Sixty-two cats (11 clinically healthy, 51 with diabetes mellitus) were included in the study. The concentration of β-hydroxybuyrate (β-HB) was measured in venous and capillary blood with a hand-held ketone meter (Precision Xceed; assay range 0-8 mmol/L) and compared with a spectrophotometric method. Precision, accuracy, and the effects of hematocrit and anticoagulants were evaluated.

    Between-run precision using low- and high-concentration control solutions was 8.1% and 2.6%, respectively; within-run coefficient of variation determined using 12 feline blood samples was 2.8%. In the 62 cats, β-HB concentrations measured with the portable ketone meter ranged from 0-7.4 mmol/L (median 0.9 mmol/L). When β-HB concentrations measured by the portable meter were < 4.0 mmol/L there was good agreement with the reference method, but concentrations > 4.0 mmol/L were lower than those obtained by the reference method in 20 of 24 cats (83%). There was good correlation between capillary and venous measurements. Results were not affected by hematocrits from 0.17 to 0.50 L/L, but EDTA was not a suitable anticoagulant.

    Measurement of β-HB concentration in peripheral or capillary blood by an easy-to-use portable ketone meter was suitable for detecting ketonemia in cats. Underestimation of β-HB concentration was observed at higher values, but results were sufficiently high to aid in diagnosing diabetic ketoacidosis.