    Evaluation of three different point-of-care tests for quantitative measurement of canine C-reactive protein (2015)

    Art
    Zeitschriftenartikel / wissenschaftlicher Beitrag
    Autoren
    Jasensky, Anne-Katherine
    Klenner, Stefanie
    Einspanier, Ralf
    Kohn, Barbara (WE 20)
    Quelle
    Veterinary clinical pathology; 44(2) — S. 205–214
    ISSN: 0275-6382
    Sprache
    Englisch
    Verweise
    DOI: 10.1111/vcp.12264
    Pubmed: 25962332
    Kontakt
    Klinik für kleine Haustiere

    Oertzenweg 19 b
    Haus 1
    14163 Berlin
    Tel.+49 30 838 62356 Fax: +49 30 - 838 460 157
    email: kleintierklinik@vetmed.fu-berlin.de

    Abstract / Zusammenfassung

    C-reactive protein (CRP) is a major acute phase protein in both people and dogs.

    The objective of this study was to evaluate the precision and accuracy of 3 different point-of-care tests (POCT) for canine CRP (cCRP) in comparison to a standard ELISA test, and to assess storage stability.

    Blood samples were collected from 125 dogs (23 healthy and 102 diseased). Serum cCRP measurements were performed using the TECOmedical AG, the EUROLyser Diagnostica, and the LifeAssays POCT. The TECOmedical AG ELISA validated for canine serum was used as a reference method. Statistical analyses included descriptive statistics, inter assay variance for each POCT, comparison of results from each POCT with the ELISA using Spearman's correlation and Bland-Altman plots, and a t-test to evaluate sample stability.

    ELISA cCRP values ranged from 0.1-4.7 mg/L (median 1.3) in healthy dogs and from 0-282 mg/L (median 17.3) in diseased dogs. Spearman's correlation coefficient between ELISA and the respective POCT measurements for cCRP concentration was 0.89 for TECOmedical, 0.85 for EUROLyser, and 0.97 for LifeAssays. Bias calculated from Bland-Altman difference plots was 27.6% for TECOmedical, -14.2% for EUROLyser, and -15.7% for LifeAssays. Inter assay reliability was > 0.9 for all POCT. Total observed error of the EUROLyser was 28.2% and therefore met the acceptable total error of 29.6%.

    All POCTs were able to measure cCRP, but precision, accuracy, and correlation coefficients varied among the 3 systems. Therefore, serial measurements for monitoring of cCRP in dogs should always be performed using the same POCT system.