    Evaluation of a novel feline AB blood typing device (2014)

    Art
    Zeitschriftenartikel / wissenschaftlicher Beitrag
    Autoren
    Hourani, Layla (WE 20)
    Weingart, Christiane (WE 20)
    Kohn, Barbara (WE 20)
    Quelle
    Journal of feline medicine and surgery; 16(10) — S. 826–831
    ISSN: 1098-612x
    Sprache
    Englisch
    Verweise
    DOI: 10.1177/1098612X14522052
    Pubmed: 24518253
    Kontakt
    Klinik für kleine Haustiere

    Oertzenweg 19 b
    Haus 1
    14163 Berlin
    Tel.+49 30 838 62356 Fax: +49 30 - 838 460 157
    email: kleintierklinik@vetmed.fu-berlin.de

    Abstract / Zusammenfassung

    This prospective study evaluated a novel immunochromatographic (IC) blood typing test for the AB blood group system. Typing was conducted comparatively on ethylenediamine tetra-acetic acid-anticoagulated blood samples from 89 sick and 16 healthy cats with the IC test, as well as two tests as reference methods, a tube agglutination and a gel column test. The samples were between 0 and 10 days old (median 3 days) and were tested for haemolysis and agglutination; the packed cell volume ranged from 0.07 to 0.57 l/l (median 0.40 l/l). The reference methods agreed with each other in 100% of the test runs. Of the 85 samples tested as blood type A by the two reference methods, 80 were correctly identified by the IC test, four were misidentified as AB and one was rated inconclusive. All B samples were correctly typed. Two of the three AB samples were correctly identified by the IC test and one was rated inconclusive. The sample quality had no influence on test performance. Of 30 repeats, 28 were readable and showed agreement in 27 cases. The agreement of the IC test with the control methods was 96.1% for the 103 conclusive tests, and it showed high sensitivity and specificity for A and B antigen detection. It is suggested that AB results be reconfirmed with a laboratory method and that a 'back-typing' be performed with plasma from B samples to detect the presence of alloantibodies. Given its very good performance and ease of use, the IC test can be recommended for clinical settings.